Resistance to electrostatic discharges. Electromagnetic compatibility of technical means. Requirements and test methods

 

The degree of rigidity of the test electrostatic discharge

1st.zh. - Air ±2kV Contact ±2kV

2st.zh. - Air ±4kV Contact ±4kV

3st.w. - Air ±8kV Contact ±6kV (Often used)

4st.w. - Air ±15kV Contact ± 8kV

Types of tests electrostatic

discharge Air discharge - the spout of the gun is round, tested with a spark

Contact - the nose of the gun is sharp tested in the butt to the product without a spark

Test method electrostatic discharge

Direct impact - interference is applied directly to the product body

Indirect impact - interference is applied through a 10x10 cm plate at a distance of 10 cm from the product

Interference is supplied by 10 pulses (after 1 second) of each polarity, air and contact discharge, each direct and indirect exposure, a total of 12 tests of 10 pulses, a total of 120 pulses. Each degree of rigidity must be confirmed by steps from the lower to the highest. If it is tested on 4 st.zh. then the total number of tests is 12 *4 =48 the total number of pulses supplied is 48*10 =480. The test time of one degree of rigidity takes about 10 minutes without taking into account the preparation.

Electrostatic discharge test schemes

1 Product with grounding, small (fits on a table of 1x0.6 m)

2. The product without grounding is small (fits on a table of 1x0.6 m)

3. The product is large (cabinet, machine), installed on a pallet 10 cm high with a grounding plate

4. The product is stationary, tests are carried out at the facility with the installation of a grounding plate

Parameters of test signals electrostatic discharge

Voltage from 2 to 15 kV

Current - front 0.7–1 ns; amplitude — 7.5; 15; 22.5; 30 A. The current decreases to 4; 8; 12; 16 A after 30 ns and another

2 times after 60 ns

Pulse power from 15kW to 450kW

Energy 25 J

The resistance of the source \ generator is 500 ohms